NIST Building 423 IAQ Manufactured House - Instrument Documentation

This document provides detailed descriptions and specifications for all active instruments in the NIST Building 423 IAQ Manufactured House Data Acquisition and Backup System.

Table of Contents

  1. Met One AIO 2 Sonic Weather Sensor
  2. Vaisala HUMICAP HMP155 Humidity and Temperature Probe
  3. Vaisala HUMICAP HMP45A Humidity and Temperature Probe

Met One AIO 2 Sonic Weather Sensor

Description: Met One AIO 2 Sonic Weather Sensor - 10m MET Tower
Manufacturer: Met One Instruments
Model: AIO 2
Status: ACTIVE
Measurement: Wind Speed, Wind Direction, Ambient Temperature, Relative Humidity, Barometric Pressure
Output: RS-232, RS-485, SDI-12
Connection: Serial (RS-232) - separate from cDAQ
Date Format: %Y%m%d
Base Path: MH DAQ/weather_station
Datetime Columns: Date, Time

Variables

  • Wind_Speed_m/s
  • Wind_Direction_deg
  • Ambient_Temperature_degC
  • Relative_Humidity_%
  • Barometric_Pressure_mb

Specifications

Parameter Range Accuracy Resolution
Wind Speed 0-75 m/s (calibrated 0-60 m/s) ±0.5 m/s or 5% of reading (whichever is greater) 0.1 m/s
Wind Direction 0-360° ±5° (including compass) 1.0°
Temperature -40°C to +60°C ±0.2°C (0-60°C), ±0.5°C (-40-0°C) 0.1°C
Relative Humidity 0-100% ±3% at 25°C 1.0%
Barometric Pressure 500-1100 hPa (mbar) ±0.5 hPa at 25°C 0.1 hPa

Notes

  • Sensor installed on a 10m MET Tower.
  • Data logged via serial output at 1 Hz.
  • Tab-delimited format with Date and Time columns.
  • NOT connected to cDAQ - uses separate serial connection.

Vaisala HUMICAP HMP155 Humidity and Temperature Probe

Description: Vaisala HUMICAP HMP155 Humidity and Temperature Probe
Manufacturer: Vaisala
Model: HMP155
Status: ACTIVE
Measurement: Relative Humidity, Temperature
Output: 0-10 VDC
Excitation: 24 VDC
DAQ Connection: Module 3-5 (NI 9201)
Date Format: %Y%m%d
Base Path: MH DAQ/indoor_daq
Datetime Columns: Date, Time

Variables

Relative Humidity:

  • RH_Bed1_M3_C0
  • RH_Liv_M3_C3
  • RH_Fam_M3_C4
  • RH_HVAC-S_M4_C3

Temperature:

  • T_Bed1_M4_C4
  • T_Liv_M4_C7
  • T_Fam_M5_C0
  • T_HVAC-S_M5_C7

Specifications

Parameter Range Accuracy Notes
Relative Humidity 0-100% RH ±1% RH (0-90% RH), ±1.7% RH (90-100% RH) at 15-25°C
Relative Humidity 0-100% RH ±(1.0 + 0.008 × reading) % RH at -20 to +40°C
Relative Humidity 0-100% RH ±(1.2 + 0.012 × reading) % RH at -40 to -20°C
Relative Humidity 0-100% RH ±(1.2 + 0.012 × reading) % RH at +40 to +60°C
Relative Humidity 0-100% RH ±(1.4 + 0.032 × reading) % RH at -60 to -40°C
Temperature -80°C to +60°C ±(0.226 - 0.0027 x reading) °C at -80 to +20°C
Temperature -80°C to +60°C ±(0.055 + 0.0057 x reading) °C at +20 to +60°C

Notes

  • Higher accuracy probes used at key monitoring locations.
  • Recommended calibration interval: 1 year.

Vaisala HUMICAP HMP45A Humidity and Temperature Probe

Description: Vaisala HUMICAP HMP45A Humidity and Temperature Probe
Manufacturer: Vaisala
Model: HMP45A
Status: ACTIVE
Measurement: Relative Humidity, Temperature
Output: 0-1 VDC
Excitation: 7-35 VDC
DAQ Connection: Module 3-5 (NI 9201)
Date Format: %Y%m%d
Base Path: MH DAQ/indoor_daq
Datetime Columns: Date, Time

Variables

Relative Humidity:

  • RH_Bed2_M3_C1
  • RH_Bed3_M3_C2
  • RH_MBa_M3_C5
  • RH_Ba1_M3_C6
  • RH_UR_M3_C7
  • RH_Gar_M4_C0
  • RH_Attic_M4_C1
  • RH_Out_M4_C2

Temperature:

  • T_Bed2_M4_C5
  • T_Bed3_M4_C6
  • T_MBa_M5_C1
  • T_Ba1_M5_C2
  • T_UR_M5_C3
  • T_Gar_M5_C4
  • T_Attic_M5_C5
  • T_Out_M5_C6

Specifications

Parameter Range Accuracy Notes
Relative Humidity 0.8-100% RH ±1% RH at 20°C factory calibration
Relative Humidity 0-100% RH ±2% RH (0-90% RH), ±3% RH (90-100% RH) field accuracy
Relative Humidity - better than 1% RH/year long term stability
Temperature -39.2°C to +60°C ±0.2°C at 20°C

Notes

  • Legacy sensors - planned replacement with HMP155.
  • Used at secondary monitoring locations.