UT181A Protocol: Reverse-Engineered Specification

Based on cross-referencing three independent community implementations:

All three implementations agree on every protocol detail documented here.

Confidence levels:

  • [KNOWN] -- confirmed by 3 independent implementations or official manual
  • [VENDOR] -- read out of the vendor application UT181A.exe (V1.05); see reverse-engineering-approach.md
  • [DEDUCED] -- logical inference
  • [UNVERIFIED] -- needs device testing

[KNOWN] here means the three community implementations agree, which is not the same as tested on a meter. What a real UT181A has confirmed, and what is still outstanding, is tracked in docs/verification-backlog.md.


1. Device Overview -- [KNOWN]

Parameter Value
Display 60,000 counts, 4-5/6 digits
Screen 3.5" 64K color TFT LCD (320x240)
Safety CAT IV 600V / CAT III 1000V
DCV accuracy 0.025% + 5 counts
Chipset Cyrustek ES51997 analog frontend
MCU STM32F103
USB bridge CP2110 (VID 0x10C4, PID 0xEA80) on older units; CH9329 (VID 0x1A86, PID 0xE429) on current production
Sample rate 2 Sa/s (60K counts), 10 Sa/s (600 counts)
Data logging 20,000 saved measurements
Recording Up to 20 named recordings
Battery 7.4V 2200 mAh Li-ion + CR2032 backup

Prerequisite: User must enable communication on the meter before each session: SETUP -> Communication -> ON. This setting resets on power cycle.


2. Transport Layer -- [KNOWN]

2.1 USB Configuration

Parameter Value
USB VID 0x10C4 (Silicon Labs), or 0x1A86 (WCH) on a CH9329 cable
USB PID 0xEA80 (CP2110 default), or 0xE429 (CH9329)
Baud rate 9600
Data bits 8
Parity None
Stop bits 1
Flow control None

Older units use the same CP2110 bridge as the UT61E+ and UT8803, with the standard CP2110 UART enable + configure. Current production ships a UT-D09 built on a WCH CH9329 instead; the UART framing above is identical either way, only the USB transport differs.

2.2 Header Byte Clarification

Important: The UT181A wire bytes are 0xAB, 0xCD -- identical to the UT61E+. The "reversed 0xCDAB header" description found in some sources (including the antage Protocol.md) refers to reading these two bytes as a little-endian uint16: byte[0]=0xAB, byte[1]=0xCD → LE uint16 = 0xCDAB. The UT61E+ reads the same bytes as big-endian: 0xABCD. The wire bytes are the same; only the host-side integer interpretation differs.


3. Frame Format -- [KNOWN]

+------+------+--------+--------+-----------+--------+--------+
| 0xAB | 0xCD | len_lo | len_hi | payload   | chk_lo | chk_hi |
+------+------+--------+--------+-----------+--------+--------+
 byte 0 byte 1 byte 2   byte 3   bytes 4..   last 2 bytes
Field Size Encoding Description
Magic 2 Fixed 0xAB 0xCD
Length 2 uint16 LE payload_size + 2 (includes checksum bytes)
Payload N Variable Command or response data
Checksum 2 uint16 LE Sum of all bytes from offset 2 through end of payload

Checksum Algorithm

checksum = sum of bytes[2] through bytes[3 + payload_size - 1]
         = length_lo + length_hi + payload[0] + ... + payload[N-1]

The checksum covers the length field and all payload bytes. It does not include the 2-byte magic header.

[VENDOR] confirmation. The vendor app builds every outgoing frame in one generic sender at 0x870408, taking (connection, opcode, payload_ptr, payload_len). It allocates payload_len + 7, writes 0xAB 0xCD, then payload_len + 3 as a uint16 LE length, then the opcode byte, then the payload, then a uint16 LE sum of every byte from offset 2 through the end of the payload. That is byte-for-byte the build_command helper in crates/dmm-lib/src/protocol/ut181a/command.rs. Every command below is a thin wrapper that fills a small stack buffer and calls it.

Key Differences from UT61E+

Aspect UT61E+ UT181A
Header bytes 0xAB 0xCD 0xAB 0xCD (same)
Length field 1 byte 2 bytes (uint16 LE)
Length meaning Bytes after length (payload + checksum) Payload + 2 (payload + checksum size)
Checksum 16-bit BE sum of all bytes before checksum 16-bit LE sum of length + payload only
Values 7-byte ASCII display string float32 LE (IEEE 754)
Communication Polled (request/response) Monitor mode (streaming) + commands

4. Packet Types -- [KNOWN]

4.1 Response Types (Device -> Host)

Code Type Description
0x01 Reply Code OK (0x4F4B = "OK") or Error (0x4552 = "ER")
0x02 Measurement Real-time measurement data
0x03 Save Saved measurement with timestamp
0x04 Record Info Recording metadata (name, interval, stats)
0x05 Record Data Recording samples (batched)
0x72 Reply Data Generic data reply (e.g., saved/recording counts)

4.2 Commands (Host -> Device)

Code Command Parameters Description
0x01 SET_MODE uint16 LE mode word Set measurement mode. [VENDOR] — wrapper at 0x8702b8 writes [lo, hi] and calls the sender with opcode 0x01; see §6.1
0x02 SET_RANGE uint8 (0x00 = auto, else 1-based) Set range. [VENDOR] — wrapper at 0x8702d8 sends is_auto ? 0 : index; see §7.1
0x03 SET_REFERENCE float32 LE Set relative reference value. [VENDOR] — wrapper at 0x87059c, driven by the REL edit box (tmrRelTimer, 0x86f2c4)
0x04 SET_MIN_MAX uint8 (0 or 1) Enable/disable min/max. [VENDOR] — wrapper at 0x870588 sends one payload byte, so the frame is 8 bytes total. antage and sigrok describe a uint32; the vendor app disagrees
0x05 SET_MONITOR uint8 (0 or 1) Enable/disable streaming. [VENDOR] — wrapper at 0x870344
0x06 SAVE_MEAS (none) Save current measurement. [VENDOR] — wrapper at 0x8703fc, zero-length payload, called from the Max/Min dialog's Save button
0x07 GET_SAVED_MEAS uint16 LE index (1-based) Retrieve saved measurement
0x08 GET_SAVED_COUNT (none) Get count of saved measurements
0x09 DEL_SAVED_MEAS uint16 LE index (0xFFFF = all) Delete saved measurement(s)
0x0A START_RECORDING name(11) + interval(2) + duration(4) Start recording
0x0B STOP_RECORDING (none) Stop recording
0x0C GET_REC_INFO uint16 LE index (1-based) Get recording metadata
0x0D GET_REC_SAMPLES uint16 LE index + uint32 LE offset(1-based) Get recording data. [VENDOR] — wrapper at 0x8703a4, 6-byte payload
0x0E GET_REC_COUNT (none) Get count of recordings
0x0F DEL_RECORDING uint16 LE index Delete recording [VENDOR] — confirmed from UT181A.exe decompilation: called after "Are you sure that you want to delete this record?" dialog, followed by GET_REC_COUNT refresh. Not in community implementations.
0x12 HOLD (button press) 0x5A = HOLD button code [VENDOR] — wrapper at 0x8703e8 hard-codes a single payload byte 0x5A (mov BYTE PTR [esp],0x5a) and is the Hold action's only call. This matches antage's toggle_hold, the only community implementation that transmits 0x12; sigrok defines the opcode but never sends it. Whether bare [0x12] also works is untested.

5. Measurement Packet (Type 0x02) -- [KNOWN]

5.1 Common Header (5 bytes)

Offset Size Field Description
0 1 misc Bit flags (see below)
1 1 misc2 Bit flags (see below)
2 2 mode uint16 LE mode word
4 1 range 0x00 = auto, 0x01-0x08 = manual

misc byte:

Bit Mask Meaning
1 0x02 Has aux1 display value
2 0x04 Has aux2 display value
3 0x08 Has bargraph / fast mode
4-6 0x70 Format: 0x00=normal, 0x10=relative, 0x20=min/max, 0x40=peak
7 0x80 HOLD active

misc2 byte:

Bit Mask Meaning
0 0x01 Auto-range active
1 0x02 High voltage warning
3 0x08 Lead error
4 0x10 COMP (comparator) mode active
5 0x20 Record mode active

5.2 Value Encoding

Full value (13 bytes):

Offset Size Field
0 4 float32 LE (IEEE 754)
4 1 Precision byte
5 8 Unit string (null-terminated)

Short value (5 bytes, used in min/max sub-values):

Offset Size Field
0 4 float32 LE
4 1 Precision byte

Precision byte:

Bits Meaning
0 Positive overload (OL)
1 Negative overload (-OL)
4-7 Decimal places (0-15)

5.3 Measurement Variants

Normal (format 0x00) -- after 5-byte header:

  • Main value: 13 bytes (float32 + precision + unit)
  • Aux1: 13 bytes (optional, if misc bit 1 set)
  • Aux2: 13 bytes (optional, if misc bit 2 set)
  • Bargraph: float32 + 8-byte unit (optional, if misc bit 3 set) -- 12 bytes, no precision byte

Hardware-confirmed 2026-09-02 (issue #5): a V AC frame with all three optional fields present has a 57-byte payload, which only accounts as 6 + 13 + 13 + 13 + 12. The bargraph's missing precision byte is what makes the arithmetic close. What its float32 means is still open -- it read 241.02 VAC against a 239.22 VAC main reading in the same frame, so it is not the displayed value.

Relative (format 0x10):

  • Relative value: 13 bytes
  • Reference value: 13 bytes
  • Absolute value: 13 bytes
  • Fast value: conditional on misc bit 3

Min/Max (format 0x20):

  • Current: 5 bytes (short value)
  • Max: 5 bytes + uint32 LE timestamp (seconds from start)
  • Average: 5 bytes + uint32 LE timestamp
  • Min: 5 bytes + uint32 LE timestamp
  • Unit: 8 bytes (shared)

Peak (format 0x40):

  • Max: 13 bytes (full value with unit)
  • Min: 13 bytes (full value with unit)

5.4 COMP Mode Extension

When misc2 bit 4 (COMP) is set, after the bargraph unit field:

Offset Size Field
0 1 Comparison mode: 0=INNER, 1=OUTER, 2=BELOW, 3=ABOVE
1 1 Result: 0=PASS, 1=FAIL
2 1 Precision/digits — low nibble, unshifted (sigrok protocol.c:112: "1 byte digits, not shifted as in other precision fields")
3 4 High limit (float32 LE)
7 4 Low limit (float32 LE, only for INNER/OUTER modes)

6. Mode Word Table -- [KNOWN]

The mode word is uint16 LE with structured nibble encoding:

  • Nibble 3 (MSB): measurement function family
  • Nibble 2: sub-function
  • Nibble 1: variant (1=normal, 2=Hz/peak/ACDC, 3=peak, 4=LPF, etc.)
  • Nibble 0 (LSB): 1=standard, 2=REL variant

79 total modes (count corrected 2026-06: both sigrok and antage define 79). Three are hardware-confirmed (marked ✓): 0x3111 (2026-04-07), 0x4211 and 0x1121 (2026-09-02, issue #5). Selected examples:

Mode Code Description
V AC 0x1111 V AC
V AC REL 0x1112 V AC relative
V AC Hz 0x1121 V AC frequency ✓ (aux1 = Hz, aux2 = period)
V AC Peak 0x1131 V AC peak
V AC LPF 0x1141 V AC low-pass filter
V AC dBV 0x1151 V AC dBV
V AC dBm 0x1161 V AC dBm
mV AC 0x2111 mV AC
mV AC+DC 0x2141 mV AC+DC coupled
V DC 0x3111 V DC ✓
V DC AC+DC 0x3121 V DC AC+DC coupled
V DC Peak 0x3131 V DC peak
mV DC 0x4111 mV DC
Temp C T1(T2) 0x4211 Temperature C, T1 main, T2 aux ✓
Temp C T2(T1) 0x4221 Temperature C, T2 main, T1 aux
Temp C T1-T2 0x4231 Temperature C, differential
Temp F T1(T2) 0x4311 Temperature F, T1 main
Resistance 0x5111 Resistance
Continuity 0x5211 Continuity (short)
Conductance 0x5311 Conductance (nS)
Diode 0x6111 Diode test
Capacitance 0x6211 Capacitance
Frequency 0x7111 Frequency
Duty Cycle 0x7211 Duty cycle
Pulse Width 0x7311 Pulse width
uA DC 0x8111 uA DC
uA AC 0x8211 uA AC
mA DC 0x9111 mA DC
mA AC 0x9211 mA AC
A DC 0xA111 A DC
A AC 0xA211 A AC

Each mode has REL variant (+1 to LSB nibble), and current/voltage modes have Hz, Peak, and AC+DC variants.

6.1 Mode switching (SET_MODE) -- [VENDOR]

Traced out of the Setting dialog (TfrmSetting) of the vendor application, V1.05. See reverse-engineering-approach.md, "Phase 3", for how the handlers were recovered and how to reproduce this. No command described here has been sent to a meter. Some of the mode words below have been observed coming from one (§6); none has been set from the host.

Composition rule

The dialog composes the word it sends from three Delphi Tag properties (0x86d700):

word = ActivePage.Tag                 ; dial family, high 12 bits
     + (primary_radio.Tag   << 4)     ; nibble 1
     +  secondary_radio.Tag           ; nibble 0

ActivePage is the tab sheet for the dial position; the primary and secondary radios live in the group boxes captioned "Primary Mode" (Tag = 1) and "Secondary Mode" (Tag = 2).

The receive path decomposes the same word (0x86cfc0, called with the mode field of every measurement packet): it makes visible only the tab whose Tag equals word & 0xFF00, then checks the primary radio whose Tag equals (word & 0xF0) >> 4 and the secondary radio whose Tag equals word & 0x0F. Send side and receive side therefore agree on the nibble layout already described in §6 — this is a direct vendor confirmation of it, not an inference.

Two consequences for a host implementation:

  • The PC can only move the meter inside the family the dial has selected. The vendor app hides every other tab, so it never emits a word with a different high byte. Turning the dial stays the user's job.
  • The app suppresses the send when the composed word equals the last word received, and sleeps 100 ms after each SET_MODE (btnUpdate1Click, 0x86ceac: call 0x8702b8 then push 0x64; call Sleep).

Dial families and primary variants

Base words are assigned in FormCreate (0x86d31c) as PageControl1.Pages[i].Tag, in tab order. Captions below are the TRadioButton.Caption values from the form resource, or the caption the click handler writes at runtime where it overrides the resource.

Family (base) n1=1 n1=2 n1=3 n1=4 n1=5 n1=6 REL for n1
V AC 0x1100 VAC VAC,HZ Peak LowPass dBV dBm 1, 4, 5, 6
mV AC 0x2100 mVAC mVAC,HZ Peak AC+DC -- -- 1, 4
V DC 0x3100 VDC AC+DC Peak -- -- -- 1, 2
mV DC 0x4100 mVDC Peak -- -- -- -- 1
Celsius 0x4200 T1,T2 T2,T1 T1-T2 T2-T1 -- -- 1, 2
Fahrenheit 0x4300 T1,T2 T2,T1 T1-T2 T2-T1 -- -- 1, 2
Ohm 0x5100 OHM -- -- -- -- -- 1
Beeper 0x5200 Beeper -- -- -- -- -- none
ns 0x5300 ns -- -- -- -- -- 1
Diode 0x6100 Diode -- -- -- -- -- none
Cap 0x6200 Cap -- -- -- -- -- 1
Hz 0x7100 Hz -- -- -- -- -- 1
Duty 0x7200 % -- -- -- -- -- 1
ms-Pulse 0x7300 ms-Pulse -- -- -- -- -- 1
uA DC 0x8100 uADC AC+DC Peak -- -- -- 1, 2
uA AC 0x8200 uAAC uAAC,Hz Peak -- -- -- 1
mA DC 0x9100 mADC AC+DC Peak -- -- -- 1, 2
mA AC 0x9200 mAAC mAAC,Hz Peak -- -- -- 1
A DC 0xA100 ADC AC+DC Peak -- -- -- 1, 2
A AC 0xA200 AAC AAC,Hz Peak -- -- -- 1

The mode word for a cell is base + (n1 << 4) + n0. For example V AC dBm plain is 0x1100 + (6 << 4) + 1 = 0x1161, and V DC AC+DC relative is 0x3100 + (2 << 4) + 2 = 0x3122.

Secondary nibble

n0 Meaning
1 Plain — the primary variant with no modifier. The radio's caption just repeats the primary's
2 REL, where the primary variant offers it (right-hand column above). Beeper and Diode instead use it for Open and Alarm
3 Unreachable — see below

A third secondary radio captioned "Peak" (Tag = 3) exists on every tab, but it is Visible = False in the form resource on eight of them and is disabled by the primary click handlers on the rest, so no n0 = 3 word is reachable from the vendor UI.

REL gating is done by the primary radio's click handler, which calls TControl.SetEnabled (0x4aeb9c) and SetCaption (0x4aecdc) on the three secondary radios. The "REL for n1" column above is that gating, read out of each handler: REL is offered on the plain, AC+DC, LowPass, dBV, dBm and T1,T2 / T2,T1 variants, and withheld on every Hz variant, every Peak variant and the two differential-temperature variants (T1-T2, T2-T1). Families with a single primary radio (Ohm, Beeper, ns, Diode, Cap, Hz, Duty, ms-Pulse) have no such handler; their secondary group is what the form resource declares.

Caveats and vendor quirks

  • mV AC+DC (0x2141) — [UNVERIFIED]. The AC+DC radio on the mVAC tab carries Tag = 4, so the vendor app does emit 0x2141, agreeing with the community row in §6. Two things undercut it: the radio is literally named rbtnmVDC_M2 with handler rbtnmVDC_M2Click (0x86e584) — a copy-paste from the mVDC tab — and the receive-side label decoder FUN_0085e69c has no 0x40 case for family 0x21, so a meter reporting 0x2141 would show a blank secondary label in the vendor app's own record grid. Whether the meter accepts the word needs hardware.
  • Duty tab. rbtnDuty_M1 has no OnClick at all in the form resource, so clicking it never refreshes the cached primary nibble. In practice the nibble is already 1 (set from the last received word by 0x86cfc0), so 0x7211 / 0x7212 still come out right.
  • btnUpdate1Click guard. Before composing, the handler returns early if the live family is 0x1100 and both rbtnVAC_M6 (dBm, field +0x430) and rbtnVAC_F3 (Peak, field +0x440) are checked — the app refuses to ask for V AC dBm + Peak. Since F3 is disabled everywhere it can be reached, this path is normally dead code. Field offsets read from the TfrmSetting published-field RTTI table.
  • Independent corroboration. The main window enables three mode-specific buttons on (word & 0xFFF0) == 0x1140 (LowPass), == 0x5210 (Beeper) and == 0x6110 (Diode), which exercises the same nibble split from a completely different code path.

Agreement with the community table (§6)

FUN_0085e69c builds the "Pri" / "Sec" labels of the record grid by switching on the mode word's high byte and on low & 0xF0. Its cases settle several rows that §6 previously carried on community agreement alone:

Word Vendor label Note
0x1121 VAC,Hz Also hardware-confirmed (§6)
0x1141 / 0x1151 / 0x1161 LowPass / dBV / dBm
0x3121 AC+DC Not Hz
0x4121 Peak Confirms mV DC Peak is 0x4121; sigrok's alternative 0x4131 is not what the vendor uses
0x4211 / 0x4221 / 0x4231 / 0x4241 T1,T2 / T2,T1 / T1-T2 / T2-T1 0x43x1 identical for Fahrenheit
0x8121 / 0x9121 / 0xA121 AC+DC Confirms the 2026-06 correction: DC-current n1=2 is AC+DC, not Hz
0x8221 / 0x9221 / 0xA221 uAAC,Hz / mAAC,Hz / AAC,Hz AC-current n1=2 is Hz
0x5212 / 0x6112 Beeper Open / Diode Alarm n0=2 here is not REL

Families 0x51, 0x52, 0x53, 0x61, 0x62, 0x71, 0x72, 0x73 have a single label each with no n1 switch, matching their single primary radio.


7. Range Byte -- [KNOWN]

Value mV V uA mA A Ohm Hz Cap
0x00 Auto Auto Auto Auto Auto Auto Auto Auto
0x01 60mV 6V 600uA 60mA -- 600R 60Hz 6nF
0x02 600mV 60V 6000uA 600mA -- 6kR 600Hz 60nF
0x03 -- 600V -- -- -- 60kR 6kHz 600nF
0x04 -- 1000V -- -- -- 600kR 60kHz 6uF
0x05 -- -- -- -- -- 6MR 600kHz 60uF
0x06 -- -- -- -- -- 60MR 6MHz 600uF
0x07 -- -- -- -- -- -- 60MHz 6mF
0x08 -- -- -- -- -- -- -- 60mF

Temperature: fixed range. Current A: fixed at 10A.

7.1 Vendor range ladders -- [VENDOR]

The Setting dialog gives each dial family one "Range" combo box, inside the group box tagged 3. Item 0 is always Auto. cbBoxRangeChange (0x86cf0c) passes the combo's ItemIndex and ItemIndex == 0 straight to the SET_RANGE wrapper (0x8702d8), which sends is_auto ? 0 : index as a single byte; the receive path does the exact inverse, writing the meter's range byte into the combo's ItemIndex (0x86d1c4). A manual range is therefore a 1-based index into the ladder below, and it is the same numbering the meter reports back in the measurement packet's range field.

Combo contents. The resource stores each item as a span string ('0 - 600', '-60 - 60'); the table condenses those to the upper bound, with ± where the span is bipolar. The strings carry no unit — the unit is the dial family's.

Dial family Items after Auto (index 1, 2, 3, ...)
V AC 6 / 60 / 600 / 1000
V DC ±6 / ±60 / ±600 / ±1000
mV AC 60 / 600
mV DC ±60 / ±600
Ohm 600 / 6000 / 60000 / 600000 / 6000000 / 60000000
Cap 6 / 60 / 600 / 6000 / 60000 / 600000 / 6000000 / 60000000
Hz 60 / 600 / 6000 / 60000 / 600000 / 6000000 / 60000000
Duty 60 / 600 / 6000 / 60000
ms-Pulse 60 / 600 / 6000 / 60000
uA DC ±600 / ±6000
uA AC 600 / 6000
mA DC ±60 / ±600
mA AC 60 / 600

Cross-check against §7: the V, mV, Ohm, uA, mA, Hz and Cap ladders agree with the community table in both length and step, index for index. (The Cap numbers only line up if read as nF — 6000000 is §7's 6mF entry — which is consistent with §8 listing nF as the wire unit.) Duty and ms-Pulse are new — §7 has no column for them. The ± spans mark the DC families' bipolar ranges; they do not imply a separate range code.

Families with no manual range. The Range group box is Visible = False on A DC, A AC, Celsius, Fahrenheit, Beeper, ns and Diode, so the vendor app never sends SET_RANGE for them. This matches §7's "Temperature: fixed range" and its empty A column, and adds Beeper, ns and Diode. Those hidden combos still hold placeholder item lists copied from another family (A DC / A AC hold the mA DC list; Celsius, Fahrenheit hold the V AC list; Beeper, ns and Diode hold the Ohm list) — they are dead UI, not range data.

Two combos (Cap, mA AC) ship without an ItemIndex property in the resource, so they start unselected rather than on Auto. Cosmetic.


8. Unit Strings -- [KNOWN]

The UT181A sends unit strings as part of measurement packets (8 bytes, null-terminated). The device determines the unit, not the host.

Wire String Unit Notes
mVDC millivolt DC
VDC volt DC
mVAC millivolt AC
VAC volt AC Hardware-confirmed 2026-09-02
mVac+dc millivolt AC+DC
Vac+dc volt AC+DC
uADC microampere DC
mADC milliampere DC
ADC ampere DC
uAAC microampere AC
mAAC milliampere AC
AAC ampere AC
uAac+dc microampere AC+DC
mAac+dc milliampere AC+DC
Aac+dc ampere AC+DC
~ ohm Tilde represents omega
k~ kilohm
M~ megohm
nS nanosiemens Conductance
nF nanofarad
uF microfarad
mF millifarad
Hz hertz Hardware-confirmed 2026-09-02
kHz kilohertz
MHz megahertz
% percent Duty cycle
ms millisecond Pulse width; also the period on V AC Hz (hardware-confirmed 2026-09-02)
dBV decibel-volt
dBm decibel-milliwatt
\xB0C degrees Celsius 0xB0 = degree symbol (Latin-1); hardware-confirmed 2026-09-02
\xB0F degrees Fahrenheit 0xB0 = degree symbol (Latin-1)

9. Timestamp Format -- [KNOWN]

Used in saved measurements (type 0x03), recording info (type 0x04), and recording data (type 0x05). Packed into 32 bits:

Bits [5:0]   -> year - 2000    (range: 2000-2063)
Bits [9:6]   -> month           (1-12)
Bits [14:10] -> day             (1-31)
Bits [19:15] -> hour            (0-23)
Bits [25:20] -> minute          (0-59)
Bits [31:26] -> second          (0-59)

10. Recording Protocol -- [KNOWN]

10.1 Start Recording (Command 0x0A)

Field Size Description
Name 11 Null-terminated ASCII (max 10 chars)
Interval 2 uint16 LE, seconds (1-3600)
Duration 4 uint32 LE, seconds (up to 143,999 minutes)

Maximum 20 named recordings on the device.

[VENDOR] The wrapper at 0x8705b0 builds exactly this 17-byte payload (push 0x11, opcode 0x0A): it copies the name, forces a NUL at payload offset 9, then writes the interval at offset 11 and the duration at offset 13. The forced NUL means the vendor app caps the name at 9 characters, not the 10 the field width allows. Whether the meter itself accepts a 10-character name is untested.

10.2 Recording Info (Response Type 0x04)

Offset Size Field
0 11 Name (null-terminated)
11 8 Unit string
19 2 Interval (uint16 LE, seconds)
21 4 Duration (uint32 LE, seconds)
25 4 Sample count (uint32 LE)
29 5 Max value (float32 LE + precision)
34 5 Average value (float32 LE + precision)
39 5 Min value (float32 LE + precision)
44 4 Start timestamp (packed 32-bit)

10.3 Recording Data (Response Type 0x05)

Downloaded in chunks via command 0x0D. Each response:

Field Size Description
Count 1 Number of samples in this packet (max 250)
Samples 9 * N Per sample: float32 LE (4) + precision (1) + timestamp (4)

Download loop: request samples starting at offset 1, increment by chunk size until all samples retrieved.


11. Implementation Considerations

11.1 Device Discrimination

The UT181A shares VID 0x10C4, PID 0xEA80 with UT61E+, UT8802, and UT8803. Discrimination approaches:

  1. Frame length: UT181A uses 2-byte LE length vs UT61E+ 1-byte. Send a UT61E+ measurement request and check if the response has a valid 1-byte length or if garbage arrives.
  2. Monitor mode: Send command 0x05 (SET_MONITOR, enable). If the device starts streaming type 0x02 packets, it's a UT181A.
  3. User selection: Let the user specify the device model.

The implemented algorithm uses the monitor-mode approach, with the payload length splitting a UT181A stream from a UT171 one; it is written up in docs/detection-design.md.

11.2 Communication Mode

The meter requires "Communication ON" in settings before USB works. This is a manual step on the device -- there is no USB command to enable it. The setting resets on power cycle.

The device cannot measure while charging.

11.3 Value Encoding

Unlike UT61E+ (ASCII display string) or UT8803 (BCD/raw bytes), the UT181A sends IEEE 754 float32 values. The host receives both the numeric value and its unit string, making parsing straightforward. The precision byte indicates decimal places for display formatting.

11.4 Existing Rust Implementation

The antage/cp211x_uart crate provides CP2110 UART control in Rust and could be used directly. The antage/ut181a crate provides a complete UT181A protocol library.


12. Confidence Summary

Most protocol detail is [KNOWN] -- confirmed by three independent implementations (antage/ut181a Rust, loblab/ut181a C++, sigrok C driver). No item rests on inference alone. The command payloads, mode composition and range ladders in §4.2, §6.1 and §7.1 are additionally [VENDOR]: read out of UT181A.exe V1.05, which is where the one outright correction lives (SET_MIN_MAX takes one byte, not four).

That is agreement between implementations and with the vendor binary, not hardware coverage. A real meter has so far confirmed the transport, framing, the normal-format value layout and three of the 79 mode words; the REL, MIN/MAX, Peak and COMP formats, every remote command and the recording protocol have never run against one. The reply frame (type 0x01, "OK" / "ER") in particular stays community-sourced — the vendor app's handling of it was not traced. docs/verification-backlog.md is the live list.

Aspect Status Sources
Frame format (header, length, checksum) [KNOWN] 3 implementations + UT181A.exe 0x870408
All 15 commands (0x01-0x12) [KNOWN] antage + sigrok + loblab
Command payload sizes (0x01-0x0F, 0x12) [VENDOR] UT181A.exe send wrappers
SET_MIN_MAX payload is 1 byte, not 4 [VENDOR] UT181A.exe 0x870588
Mode word nibble layout (family / primary / secondary) [VENDOR] UT181A.exe 0x86d700 + 0x86cfc0
Per-family mode variants and captions (§6.1) [VENDOR] TfrmSetting resource + click handlers
All 79 mode words [KNOWN] antage + sigrok
Range bytes 0x00-0x08 [KNOWN] 3 implementations
Range ladders per family, 1-based index (§7.1) [VENDOR] TfrmSetting range combos
Measurement packet (all 4 variants) [KNOWN] antage + sigrok
COMP mode extension [KNOWN] sigrok driver
Unit strings [KNOWN] antage + sigrok
Timestamp format [KNOWN] 3 implementations
Recording protocol [KNOWN] 3 implementations
Transport (9600/8N1 CP2110) [KNOWN] 3 implementations
Header bytes = 0xAB 0xCD (same as UT61E+) [KNOWN] 3 codebases verified
Device specs (60K counts, modes, ranges) [KNOWN] User manual