@@ -151,8 +151,8 @@ Due to the inherent safety risks, this parameter is not enabled by default. To a
151
151
To create an insulin with another concentration, first open Insulin Manager.
152
152
153
153
- Click the "+" icon at the bottom of the screen.
154
-
- Select the new expected concentration from the dropdown menu (U10, U50, U100, U200 are available).
155
-
- diluted concentrations like U10 or U50 are for children with very low basal rates.
154
+
- Select the new expected concentration from the dropdown menu (U10, U20, U50, U100, U200 are available).
155
+
- diluted concentrations like U10, U20 or U50 are for children with very low basal rates.
156
156
- U200 is more for adults with high insulin needs.
157
157
- Adjust peak time, Duration of Insulin Action (DIA), and customize the Insulin Nickname (optional).
158
158
- Use the save button at the top right of the screen.
@@ -172,13 +172,15 @@ When you start from U100 insulin, all your profile, historical data and reports
172
172
- For a tube pump, click the Prime/Fill button (for Insulin Change) within the Overview (expanded status bar view), then click the "Change" button.
173
173
174
174
1. Select the target concentration in the dropdown menu (e.g. `U200`).
175
-
1. Then select the right insulin below.
175
+
1. Then select the right insulin below. Until you confirm, the caption above its name says **Will be activated** (instead of **Currently active**).
176
176
1.**Important:** the concentration switch is only recorded within AAPS after the Profile Switch is applied to the pump.
177
177
- This step is mandatory for all Patch/Pod pumps.
178
178
- For tube pumps, wait for the Profile Switch to be applied to the pump, or within Profile Manager or **Insulin settings**.
179
179
180
180
That's all!
181
181
182
+
If a temporary **Profile Switch** (one with a duration) is running when you change the insulin, the new insulin is also given to the profile that comes back when the temporary switch ends. The old insulin does not come back.
183
+
182
184
### Switch concentration for users with U200 insulin associated with a 50% profile (AAPS 4)
183
185
184
186
When you start from U200 already used in your pump, but associated with a 50% profile, you have to do additional steps compared to the standard switch from U100 insulin.
Copy file name to clipboardExpand all lines: docs/EN/CompatiblePumps/Accu-Chek-Combo-Pump-v2.md
+2Lines changed: 2 additions & 0 deletions
Display the source diff
Display the rich diff
Original file line number
Diff line number
Diff line change
@@ -36,6 +36,8 @@ Be aware of the following limitations:
36
36
37
37
* Extended bolus and multiwave bolus are currently not supported (you can use [Extended Carbs](../DailyLifeWithAaps/ExtendedCarbs.md) instead).
38
38
* Only one basal profile (the first one) is supported.
39
+
* The lowest basal rate in your **Profile** is 0.05 U/h. **AAPS** refuses a **Profile** with a lower rate. The Combo rounds basal rates to its own steps (0.01 U/h below 1 U/h, 0.05 U/h up to 10 U/h, 0.1 U/h above). **AAPS** uses the rates the pump really gives.
40
+
* Temporary basal rates are limited to 500% of the current basal rate. **AAPS** takes this limit into account.
39
41
* The loop is disabled if the currently active profile on the pump isn't profile no. 1. This continues until profile no. 1 is made the active one; when that is done, the next time AAPS connects (either on its own after a while or because you tap **Refresh** on the Accu-Chek Combo pump screen), it will notice that profile no. 1 is the current one, and enable the loop again.
40
42
* If the loop requests a running TBR to be cancelled, the Combo will set a TBR of 90% or 110% for 15 minutes instead. This is because actually cancelling a TBR causes an alert on the pump which causes a lot of vibrations, and these vibrations cannot be disabled.
41
43
* Bluetooth connection stability varies with different phones, causing "pump unreachable" alerts, where no connection to the pump is established anymore. If that error occurs, make sure Bluetooth is enabled, tap **Refresh** on the pump screen (**Manage** > **Pump**) to see if this was caused by an intermitted issue and if still no connection is established, reboot the phone which should usually fix this.
Copy file name to clipboardExpand all lines: docs/EN/CompatiblePumps/CareLevo.md
+7-1Lines changed: 7 additions & 1 deletion
Display the source diff
Display the rich diff
Original file line number
Diff line number
Diff line change
@@ -7,7 +7,7 @@ The CareLevo is a tubeless patch pump made by CareMedi. Each patch holds up to 3
7
7
The CareLevo driver is new in **AAPS**. Read this whole page before you activate your first patch, and keep the pump's own instructions for use at hand.
8
8
```
9
9
10
-
The smallest dose the patch can deliver is 0.05 U. Set your basal rates in the **Profile** to a minimum of 0.05 U/h in steps of 0.05 U/h, otherwise the insulin the **Profile** expects and the insulin the patch delivers differ slightly.
10
+
The smallest dose the patch can deliver is 0.05 U. Set your basal rates in the **Profile** to a minimum of 0.05 U/h in steps of 0.05 U/h, otherwise the insulin the **Profile** expects and the insulin the patch delivers differ slightly.**AAPS** uses the basal rates the patch accepted.
11
11
12
12
```{contents} Table of contents
13
13
:depth: 1
@@ -105,6 +105,12 @@ If the insertion fails, **AAPS** tells you how many attempts are left. When no a
105
105
106
106
Insulin delivery starts as soon as the insertion is confirmed.
107
107
108
+
## Bolus messages
109
+
110
+
If the patch refuses a bolus, **AAPS** tells you why. For example *Bolus can't be delivered. Bolus available: 2.8 U* means the patch can deliver only that amount now.
111
+
112
+
If you stop a bolus and the patch does not confirm the stop, **AAPS** waits for the bolus to end and records the amount the patch reports. If a bolus was not recorded (for example because it was stopped while it was running), **AAPS** adds it from the patch total and asks you to check IOB before you try again.
113
+
108
114
## Suspending and resuming insulin delivery
109
115
110
116
Press **Suspend Infusion** on the pump screen and select how long delivery should pause: from 30 minutes to 4 hours in 30-minute steps. Delivery does not resume by itself: when the pause is over, the patch raises a **Start Insulin** alert. Press **Resume Infusion** on the pump screen, or **Start Insulin** on the alert, and confirm.
Copy file name to clipboardExpand all lines: docs/EN/CompatiblePumps/DanaR-Insulin-Pump.md
+2-2Lines changed: 2 additions & 2 deletions
Display the source diff
Display the rich diff
Original file line number
Diff line number
Diff line change
@@ -24,13 +24,13 @@ _These instructions are for configuring the app and your pump if you have a Dana
24
24
* Open the pump screen: tap **Open plugin** on the DanaR card (or go to **Manage** > **Pump**).
25
25
* Tap **Pairing**. On the pairing screen:
26
26
* Enter your pump password under **Pump password** (default password is 1234).
27
-
* Under **Select paired device**, tap your DanaR serial number. If the list is empty, tap **Open Bluetooth settings** and pair the pump with your phone first (see the step above), then tap the refresh icon.
27
+
* Under **Select paired device**, tap your DanaR serial number. If the list is empty, tap **Open Bluetooth settings** and pair the pump with your phone first (see the step above). When you come back to **AAPS**, the pump appears in the list. If it does not, tap the refresh icon.
28
28
* Tap **Pairing**. **AAPS** connects to the pump and checks the password.
29
29
* To change the driver settings, tap **Settings** on the DanaR card in **Configuration** > **Pump**:
30
30
***Bolus speed** (DanaR and DanaRv2): change the default bolus speed (12 s/U, 30 s/U or 60 s/U).
31
31
***Use extended boluses for >200%** (DanaR and DanaR Korean): enable this if you want **AAPS** to allow a basal rate above 200%. Note this means you cannot loop with high TBRs while using extended boluses for food.
32
32
* Set basal step on pump to 0.01 U/h
33
-
* Set bolus step on pump to 0.1 U/h
33
+
* Set bolus step on pump to 0.1 U. **AAPS** reads the bolus step from the pump and uses it for every bolus.
Copy file name to clipboardExpand all lines: docs/EN/CompatiblePumps/DanaRS-Insulin-Pump.md
+16-8Lines changed: 16 additions & 8 deletions
Display the source diff
Display the rich diff
Original file line number
Diff line number
Diff line change
@@ -1,12 +1,14 @@
1
-
# DanaRSand Dana-i
1
+
# DanaRS, Dana-i and Dana-i2
2
2
3
-
_These instructions are for configuring the app and your pump if you have a DanaRS from 2017 onwards or the newer Dana-i. Visit [DanaR Insulin Pump](./DanaR-Insulin-Pump.md) if you have the original DanaR instead._
3
+
_These instructions are for configuring the app and your pump if you have a DanaRS from 2017 onwards, a Dana-i or a Dana-i2. Visit [DanaR Insulin Pump](./DanaR-Insulin-Pump.md) if you have the original DanaR instead._
4
4
5
5
**New Dana RS firmware v3 can be used from AAPS version 2.7 onwards.**
6
6
7
7
**New Dana-i can be used from AAPS version 3.0 onwards.**
8
8
9
-
* In DanaRS/i pump "BASAL A" is used by the app. Existing data gets overwritten.
9
+
**Dana-i2 can be used from AAPS version 4.0 onwards.** It uses the same **Dana-i/RS** driver as the Dana-i.
10
+
11
+
* In DanaRS/i/i2 pump "BASAL A" is used by the app. Existing data gets overwritten.
10
12
11
13
```{contents} Table of contents
12
14
:depth: 1
@@ -51,7 +53,7 @@ _These instructions are for configuring the app and your pump if you have a Dana
51
53
52
54
* For DanaRS v1, **AAPS** asks for the **Pump password (v1 only)**. Enter your pump password (see [Default password](#DanaRS-Insulin-Pump-default-password)).
53
55
* For DanaRS v3, **AAPS** asks you to press OK on the pump and type the 2 sequences of numbers and letters shown on the pump. Keep the pump display on by pressing the minus button until you have finished. Then tap **Done**.
54
-
* For Dana-i, the standard Android pairing dialog appears. Enter the 6-digit number shown on the pump.
56
+
* For Dana-i and Dana-i2, the standard Android pairing dialog appears. Enter the 6-digit number shown on the pump.
55
57
56
58
* When **Pairing successful!** is shown, tap **Done**.
57
59
@@ -62,9 +64,15 @@ _These instructions are for configuring the app and your pump if you have a Dana
62
64

63
65
* In **Settings**, select **Bolus speed** to change the default bolus speed (12 s/U, 30 s/U or 60 s/U).
64
66
* Set basal step on pump to 0.01 U/h using Doctors menu (see pump user guide).
65
-
* Set bolus step on pump to 0.05 U/h using Doctors menu (see pump user guide).
67
+
* Set bolus step on pump to 0.05 U using Doctors menu (see pump user guide). **AAPS** reads the basal and bolus steps from the pump and uses them for every bolus, so a bolus always matches the step set on the pump. The pump screen shows them in the **Basal/Bolus step** row.
66
68
* Enable extended boluses on pump
67
69
70
+
### Options shown only for some models
71
+
72
+
***Auto lock**: on a Dana-i2, the **User options** screen (on the pump screen, **Manage** > **Pump**) has an **Auto lock** switch for the pump's auto lock setting. Press **Save options to pump** after changing it.
73
+
***Easy mode**: on pump models that have an easy menu mode, the pump screen shows an **Easy mode** row with **On** or **Off**.
74
+
* If a Dana-i2 refuses a bolus, **AAPS** shows the reason, for example *Pump is not primed*, *Not enough insulin in the reservoir* or *Easy sleep mode is on in the pump*.
75
+
68
76
(DanaRS-Insulin-Pump-default-password)=
69
77
70
78
### Default password
@@ -137,17 +145,17 @@ When switching to a new phone the following steps are necessary:
137
145
* Install AAPS on the new phone.
138
146
*[Import settings](../Maintenance/ExportImportSettings.md) on your new phone
139
147
140
-
### DanaRS v3, Dana-i
148
+
### DanaRS v3, Dana-i, Dana-i2
141
149
* Start pairing procedure as described [above](#DanaRS-Insulin-Pump-pairing-pump).
142
150
* Sometimes you may need to clear the pairing information in **AAPS** first. On the pump screen (**Manage** > **Pump**), tap **Unpair** and confirm **Reset pairing information?**. Then tap **Pairing** again.
143
151
144
-
## Timezone traveling with DanaRSand Dana-i pumps
152
+
## Timezone traveling with DanaRS, Dana-i and Dana-i2 pumps
145
153
146
154
For information on traveling across time zones see section [Timezone traveling with pumps](#timezone-traveling-danarv2-danars).
147
155
148
156
## Where to get help
149
157
150
-
Development of the DanaRS/Dana-i driver is done by the community on a **volunteer** basis. Before requesting help, please:
158
+
Development of the DanaRS/Dana-i/Dana-i2 driver is done by the community on a **volunteer** basis. Before requesting help, please:
151
159
152
160
1.**Read** the relevant section of this documentation to confirm how the feature is meant to work.
153
161
2.**Ask** on the *#AAPS* channel on [Discord](https://discord.gg/4fQUWHZ4Mw), or in one of the other [community channels](../GettingHelp/WhereCanIGetHelp.md).
Copy file name to clipboardExpand all lines: docs/EN/CompatiblePumps/EOPatch2.md
+3-1Lines changed: 3 additions & 1 deletion
Display the source diff
Display the rich diff
Original file line number
Diff line number
Diff line change
@@ -2,7 +2,7 @@
2
2
3
3
The patch needs rapid-acting U-100 insulin, such as NovoRapid or Humalog. Use the rapid-acting insulin your doctor has prescribed for you.
4
4
5
-
The smallest dose the patch can deliver is 0.05 U. Set every basal rate in your **Profile** to at least 0.05 U/hr, in steps of 0.05 U/hr. Otherwise the total insulin your **Profile** expects and the insulin the patch actually delivers may not match. In the same way, boluses must be at least 0.05 U.
5
+
The smallest dose the patch can deliver is 0.05 U. Set every basal rate in your **Profile** to at least 0.05 U/hr, in steps of 0.05 U/hr. Otherwise **AAPS** rounds each rate to the nearest 0.05 U/hr (never down to 0) and uses the rates the patch really gives, so the total insulin per day differs slightly from your **Profile**. In the same way, boluses must be at least 0.05 U.
6
6
7
7
```{contents} Table of contents
8
8
:depth: 1
@@ -167,6 +167,8 @@ Replace the patch when it is low on insulin, when it expires, or if it is faulty
167
167
168
168
After that, the wizard continues straight on to activating a new patch (see above). Press the back arrow if you want to stop here.
169
169
170
+
While no patch is active, **AAPS** shows the pump as suspended and records a 0 U/hr temporary basal, so no basal insulin is counted in IOB until the next patch is activated.
171
+
170
172

Copy file name to clipboardExpand all lines: docs/EN/CompatiblePumps/Equil5.3.md
+15-1Lines changed: 15 additions & 1 deletion
Display the source diff
Display the rich diff
Original file line number
Diff line number
Diff line change
@@ -10,6 +10,8 @@ These instructions are for configuring the Equil insulin pump.
10
10
## Pump capabilities with AAPS
11
11
12
12
* Tubeless patch pump that communicates with **AAPS** over your phone's native Bluetooth, without an additional communication device.
13
+
***AAPS** treats the Equil as a patch pump with a 200 U reservoir: the main screen shows the patch pump status lights, and **Prime/Fill** and **Pump battery change** are not offered. Filling, priming and changing the battery are done in the pump wizards.
14
+
* While the pump is suspended or stopped, **AAPS** records a 0 U/h temporary basal, so no basal insulin is counted in IOB.
13
15
* DST and timezone changes must be handled manually.
14
16
15
17
## Hardware and software requirements
@@ -42,6 +44,7 @@ Open the Equil settings with **Settings** on the **Equil** card, or with the gea
42
44
***Low battery alarm**: warns you when the pump battery is low.
43
45
***Low drug storage alarm**: warns you when the insulin in the reservoir is low.
44
46
***Alert tone**: how the pump alerts you: **Mute**, **Vibrate only**, **Tone only** or **Vibrate and tone**.
47
+
***Bolus speed**: how fast a bolus is given: **1.5 U/min (standard)**, **3 U/min**, **6 U/min** or **12 U/min**. A faster bolus may give more occlusion alarms.
45
48
46
49
```{admonition} Max basal and max bolus on the patch
47
50
:class: note
@@ -96,11 +99,22 @@ If you set your own pairing password (recommended for your safety), store it som
96
99
97
100
Once the pump is paired, the pump screen shows these buttons:
98
101
102
+
***Refresh**: reads the state and the history of the pump now.
99
103
***Suspend** / **Resume delivery**: stop or restart insulin delivery.
100
-
***Change reservoir**: guides you through detaching the pump, fitting a new reservoir, priming and attaching it again.
104
+
***Change reservoir**: guides you through detaching the pump, fitting a new reservoir (and, if needed, a new battery), priming and attaching it again. The first step asks *Will you also put on a new infusion set (base plate and cannula)?*
105
+
***Yes, new infusion set**: you choose the site, the cannula is primed, and **AAPS** records an insulin change and a cannula change.
106
+
***No, only reservoir or battery**: the cannula is not primed and **AAPS** records only an insulin change.
107
+
108
+
After a reservoir change the battery level shows **-** until **AAPS** reads it again from the pump.
101
109
***History**: shows the events recorded by the pump.
102
110
***Unpair device**: guides you through detaching the pump and unpairing it from **AAPS**. This stops insulin delivery, moves the pump plunger back to the bottom, and deletes the pairing.
103
111
112
+
### Alarms
113
+
114
+
* The low insulin alarm sounds once at each level (10 U, 5 U and 2 U) and when the reservoir is empty. The low battery alarm sounds once at 10% and once at 5%. Each level alarms only once until the next reservoir change.
115
+
* Pump alarms and stops are also recorded as announcements in the treatments, at the time the pump logged them, so you can see later when and why the pump stopped.
116
+
* If the reservoir runs empty during a bolus, **AAPS** records only the insulin given until then and tells you both amounts.
117
+
104
118
## Where to get help
105
119
106
120
Development of the Equil driver is done by the community on a **volunteer** basis. Before requesting help, please:
Copy file name to clipboardExpand all lines: docs/EN/CompatiblePumps/MedtronicPump.md
+1-1Lines changed: 1 addition & 1 deletion
Display the source diff
Display the rich diff
Original file line number
Diff line number
Diff line change
@@ -140,7 +140,7 @@ The RL Status dialog has two tabs:
140
140
(medtronic-actions)=
141
141
## Actions
142
142
When the Medtronic driver is used, two additional actions are available on the pump screen (**Manage** > **Pump**):
143
-
-**Wake and Tune Up** - In the event that AAPS hasn't connected to your pump for a sustained period (it should connect every 5 minutes), you can force a Tune Up. This will try to contact your pump, by searching all of the possible radio frequencies used by your pump. In the event a successful connection is made the successful frequency will be set as the default.
143
+
-**Wake and Tune Up** - In the event that AAPS hasn't connected to your pump for a sustained period (it should connect every 5 minutes), you can force a Tune Up. This will try to contact your pump, by searching all of the possible radio frequencies used by your pump. In the event a successful connection is made the successful frequency will be set as the default. While the tune-up runs, the RileyLink status on the pump screen shows it. If a tune-up fails (for example because the pump was out of range), **AAPS** tries again by itself about every 10 minutes.
144
144
-**Reset RileyLink** - If you reset your RileyLink compatible device you may need to use this action so that device can be reconfigured (frequency set, frequency type set, encoding configured).