The RL Status dialog has two tabs:
104
103
-**Settings**: Shows settings about the RileyLink compatible device: Configured Address, Connected Device, Connection Status, Connection Error and RileyLink Firmware versions. Device Type is always Medtronic Pump, Model would be your model, Serial number is configured serial number, Pump Frequency shows which frequency you use, Last Frequency is last frequency used.
105
104
-**History**: Shows communication history, items with RileyLink shows state changes for RileyLink and Medtronic shows which commands were sent to pump.
@@ -109,7 +108,6 @@ When the Medtronic driver is used, two additional actions are added to Actions T
109
108
-**Wake and Tune Up** - In the event that AAPS hasn't connected to your pump for a sustained period (it should connect every 5 minutes), you can force a Tune Up. This will try to contact your pump, by searching all of the possible radio frequencies used by your pump. In the event a successful connection is made the successful frequency will be set as the default.
110
109
-**Reset RileyLink Config** - If you reset your RileyLink compatible device you may need to use this action so that device can be reconfigured (frequency set, frequency type set, encoding configured).
111
110
112
-
113
111
## Important notes
114
112
115
113
### Special attention in NS configuration needed
@@ -155,7 +153,6 @@ There are a number of options to try and resolve connectivity issues.
155
153
2.*While* the phone is restarting restart the Rileylink device
156
154
3. Open AAPS and allow the connection to restore
157
155
158
-
159
156
### How to determine what Frequency my pump uses
160
157

161
158
@@ -164,3 +161,10 @@ On the back of the pump you will find a line detailing your model number along w
164
161
- CA - Canada (in frequency selection you need to select "US & Canada (916 MHz)")
165
162
- WW - Worldwide (in frequency selection you need to select "Worldwide (868 Mhz)")
166
163
164
+
### How to fix "pump unreachable" error?
165
+
166
+
After updating to version 3.4.x.x a "pump unreachable error appears for some Medtronic pumps. Fix is to change pump encoding type under Medtronic preferences in AAPS:
167
+
168
+
1. Open menu in upper right corner (⁝)
169
+
2. Open "Medtronic Preferences"
170
+
3. Under "Medtronic Encoding" change to "Hardware Encoding"